RMS1650 series of ultra-high temperature insulating material resistivity measurement system is mainly used to evaluate and measure the electrical properties of insulating materials.
The system is designed based on the three-electrode method and has obtained a patent for an insulation material resistivity measurement fixture (patent number: ZL2015206666054).
The experimental methods for resistance and resistivity of insulation materials at high temperatures have been standardized and developed. The system can directly measure the resistance, bulk resistivity, and leakage current of samples under high temperature, vacuum, and atmospheric conditions, as well as the curves of these parameters changing with temperature and time.
This system has been widely used in the aerospace sensor field and is the preferred system for high-temperature insulation material resistivity measurement in the industry. Therefore, it is widely used in high-efficiency scientific research institutions and enterprise units for high-temperature insulation performance research of materials.
Product spec:
Measurement temperature: RT~1650°C
Temperature ramp rate: 0-10°C/min, typical 3°C/min
Temperature control accuracy: ±0.5°C
Resistance range: 10^3Ω~10^162Ω
Measurement accuracy: depending on the measurement range of the instrument
Measurement environment: air/flow/vacuum, optional
Temperature control mode: PID precision temperature control
Measurement principle: three-ring electrode method
Test channel: single channel
Test sample: φ < 20mm, d<5mm block
Electrode material: upper electrode semi-circular type platinum electrode, lower electrode flat type platinum electrode
Data storage format: TXT text format
Data transmission: USB communication
Patent No.: ZL2015206666054
Power supply: 220V±10%, 50Hz, 2.6kW
Operating temperature: 5°C to +40C
Storage temperature: -40C to +65C
Operating humidity: up to 95% relative humidity at +40°C (non-condensing)
Equipment size: 1260x760x1727mm (LxWxH)
Weight: 200kg
Warranty: 1 year